top of page

Kidney Stone / Prostate Surgery .

DSC_1018.jpg

Testing Lithovue (disposable URS)

Profiles of Lithovue(TM)

- 7.7Fr distal tip, 9.5Fr diameter, bidirectional 270 degrees

Over guidewire insertion into the ureter

- 7.7Fr distal tip is larger than 7.0Fr (Flex-X2), which means that lithovue could not be inserted into the ureter even after ureteroscopic (7.5Fr) dilatation of UVJ (ureterovesical junctional) orifice.

- Through the ureteral access sheath: very easy (11/13Fr size is enough)

To handle the scope

- Grip: Very light.

- Irrigation: very good (40-50 cmH2O is enough for naviation, 100 cmH2O for dusting)

- Deflection angles: 270 degrees with a 1.9Fr basket, 150-200 degress with 200 um laser fiber, 110-180 degress with 365 um laser fiber

- Maneuverability and overal performance satisfaction: very good

Working channel

- 3.6 Fr size: very good

- No friction feeling at the connection point of working sheath and the body during insertion of any disposable device (very good point!!)

- No portseal: A new portseal of 3 way connector should be used

Navigation of the renal collecting system with the Lithovue: very good image quality

조회수 126회댓글 0개

최근 게시물

전체 보기
bottom of page